| Channels |
| |
0.3, 0.5, 1.0, 2.0, 5.0, 10.0 µm |
|
| Flow Rate |
| |
0.1 cfm (2.83 L/min) controlled by
internal pump |
|
| Count Modes |
| |
Concentration, totalize, audio |
|
| Counting Efficiency |
| |
50% @0.3 µm; 100% for particles > 0.45
µm (per JIS B9921:1997) |
|
| Zero Count |
| |
1 count/5 minute (JIS B9921:1997) |
|
| Coincidence Loss |
| |
5% at 2,000,000 particles per ft3 |
|
| Relative Humidity |
| |
± 7%, 20% to 90% non-condensing |
|
| Temperature |
| |
± 3 ˚C, 10 ˚C to 40 ˚C (50 ˚F to 104
˚F) |
|
| Data Storage |
| |
5000 sample records (rotating buffer)
of date, time, counts, relative humidity, temperature, sample
volumes, alarms, label |
|
| Alarms |
| |
Counts, low battery, sensor fail |
|
| Delay Time |
|
| Sample Inlet |
|
| Interface |
| |
RS-232 and RS-485 via RJ-45 |
|
|
Environmental Specifications |
| Operating: |
| |
10 ˚C to 40 ˚C (50 ˚F to 104 ˚F), 20%
to 90% relative humidity, non-condensing |
|
| Storage: |
| |
-10 ˚C to 50 ˚C (14 ˚F to 122 ˚F), up
to 90% relative humidity, non-condensing |
|
|
Mechanical & General
Specifications |
| Size |
| |
11.43 cm x 20.96 cm x 5.72 cm (4.5 in x
8.25 in x 2.25 in) |
| |
Dimensions (W x L x D) |
|
| Weight |
|
| Power |
| |
AC adapter, 12 V DC at 2.5A, 90 to 250
V AC, 50 to 60 Hz |
|
| Rechargeable Battery |
| |
NiMH, 4.8 V at 4.5 Ah; replaceable |
|
| Battery Life |
|
| Recharge Time |
|
| Calibration |
| |
PSL particles in air (NIST traceable) |
|
|