Scope: DC voltage, DC Current, AC Voltage,
AC Current, Frequency, Pulse Width + or -, Phase, Duty cycle + or -, Peak max, Peak min,
Peak min-max, Crest Factor
Transients
Minimum pulse width: 40 ns
Useful bandwidth input 1: DC to 1 MHz
(with test leads TL24)
Number of transients 40
Voltage threshold settings: 20%, 50%, 100%,
200% above or below reference signal
Reference signal: After START, the Vrms and
frequency of the signal are measured. From these data a pure sinewave is calculated as
reference for threshold setting.
Vpeak min, Vpeak max at cursor:
10 V, 25 V, 50 V, 125 V, 250 V, 500 V, 1250 V
Accuracy: ±5% of full scale
Inrush Current
Current ranges (selectable):
1 A, 5 A, 10 A, 50 A, 100 A, 500 A, 1000 A
Inrush times (selectable):
1 s, 5 s, 10 s, 50 s, 100 s, 5 min
Cursor readings
A Peak max at cursor 1 and cursor 2
Accuracy: ± 5% of full scale
Time between cursors
4 to 235 pixels (1 pixel = inrush time/250)
Accuracy: ±(0.2% + 2 pixels)
Scope, dual channel scope with measurement reading
Input impedance
Input 1:
Ranges: 1 MΩ//12 pF; with BB 120: 20pF
Accuracy: ±2 pF; with BB120 ±3 pF
Input 2:
Ranges: 1 MΩ//10 pF; with BB 120: 18pF
Accuracy: ±2 pF; with BB120 ±3 pF
Vertical
Voltage Ranges:
Ranges: 50 mV/div to 500 V/div
Accuracy: ± (1% + 2 pixels)
Vertical sensitivity, resolution:
Ranges: 5 mV/div to 500 V/div, 8 bit (256 levels)
Bandwidth input 1 (voltage):
Ranges: DC to 20 MHz at inputs, or with BB120 and VPS 100-R probe (opt); 1 MHz with TL24
Leads
Bandwidth input 2 (current):
DC to 15 kHz at inputs
10 kHz with 80i-500s Current Clamp
Trigger source:
Ranges: Input 1 or Input 2 or Automatic selection
Trigger Mode:
Automatic Connect-and ViewTM, Free Run, Single Shot.
Connect-and-ViewTM:
Advanced automatic triggering that recognizes signal patterns and automaticaly adjusts
triggering, timebase and amplitude. Automatically displays stable pictures of complex and
dynamic signals like motor drive and control signals.
Pre Trigger:
Up to 10 divisions
Measurement readings, per channel selectable:
Ranges: Volts & Amps (DC, AC, AC + DCrms, Peak max, Peak min, Peak min/max).
Frequency, Duty cycle + or =, Phase, Pulse Width + or -, Crest factor
Temperature:**
Ranges: -100.0 ºC to 400.0 ºC, (-200.0 ºF to 800.0 ºF)
Accuracy: ±(0.5% + 5 counts)
Max current, max open circuit volt:
Ranges: 0.5 mA, < 4 V (all functions above)
Memory
Number of screens: 20
Optical Isolated RS-232 Interface
To printer:
Supports HP LaserjetTM, Deskjet, Epson FX/LQ and Postscript printers with
optional PAC91 Printer Adapter Cable
To PC:
FlukeView® Power Quality Analyzer Software with PM9080 Interface Adapter
included
FlukeView Power Quality Software
Hardware requirements:
PC or 100% compatible with Windows® 95, 98, Me, 2000, NT 4.0.
* Rated 600 V CAT III
** Requires optional temperature accessory
Line voltage adapter/battery charger included
Installed battery: Rechargeable NiCd pack (4 to 6 Vdc)
Operating time: 4 hours
Charging time: 4 hours (Fluke 43B off) 12 Hours (Fluke 43B on)
Refresh Cycle: 8 to 14 hours (to keep NiCd battery capacity optimal)
Environmental
Temperature: 0 ºC to 50 ºC (32 ºF to 122 ºF)
Environmental: MIL 28800E, Type 3, Class III, Style B
Enclosure: IP51 (dust, drip water proof)
Mechanical Data
Size: (H x W x D) 232 x 115 x 50 mm (9.1 x 4.5 x 2
inches)
Weight: 1.1 kg (2.5 lbs) incl. battery pack
Safety
For measurements on 600 Vrms Category III installations,
Pollution Degree 2, in accordance with EN61010-1 (1993) (IEC1010-1), ANSI/ISA S82.01-1994,
CAN/CSA-C22.2 No. 1010.1-92 UL3111-1
Surge protection: 6 kV on input 1 and 2
Floating measurements: 600 Vrms from any terminal to ground
Warranty
3 years parts and labor on Fluke 43B, 1 year on accessories