Contact EMSCO today to purchase your Fluke Digital Multimeter
and take advantage of the Lifetime Limited Warranty on selected models

Fluke Product Home

Fluke 789 ProcessMeter™
Specifications

Product Home - Features - Specifications - Options & Accessories - Pricing
Measurement Function - Output Function - General Specifications

Measurement Function

V dc Range and Resolution
400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V

Best Accuracy (% of reading + LSD)
0.1 % + 1

V ac (true-rms) Range and Resolution
400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V

Best Accuracy (% of reading + LSD)
0.7 % + 2

mA dc Range and Resolution
30.000 mA

Best Accuracy (% of reading + LSD)
.05 % + 2

A dc Range and Resolution
1.000 A (0.440 A continuous)

Best Accuracy (% of reading + LSD)
0.2 % + 2

A ac Range and Resolution
1.000 A (0.440 A continuous)

Best Accuracy (% of reading + LSD)
1 % + 2

Resistance Range and Resolution
400.0 Ohms, 4.000 k, 40.00 k, 400.0 k, 4.0 M, 40 M

Best Accuracy (% of reading + LSD)
0.2 % + 1

Frequency (0.5 Hz to 20 kHz) Range and Resolution
199.99 Hz, 1999.9 Hz, 19.999 kHz

Best Accuracy (% of reading + LSD)
.005 % + 1

Diode Test Range and Resolution
2.000 V (shows diode voltage drop)

Best Accuracy (% of reading + LSD)
2 % + 1

Continuity Range and Resolution
Beeps for resistance < approx. 100 ohms

 

Output Function

DC Current Output
(Internal battery operation)
Range and Resolution
0.000 to 20.000 mA or 4.000 to 20.000 mA
(selectable at power-up)
Over-range to 24.000 mA

Drive Capability (% of span)
24 V compliance, or,
1,200 Ohms, @ 20 mA

Accuracy
.05 %

DC Current Simulate
(Ext. 24 Volt loop supply)
Range and Resolution
0.000 to 20.000 mA or 4.000 to 20.000 mA,
(selectable at power-up)
Over-range to 24.000 mA

Drive Capability (% of span)
1000 Ohms, @ 20 mA

Accuracy
.05 %

24 V Loop Supply Range and Resolution
Minimum 24 V

Drive Capability (% of span)
250 Ohms @ 20 mA

Accuracy
> 24 V

Current Adjustment Modes   Range and Resolution
Manual: 0.001 mA, 0.1 mA, 25 % and 100 % step
Automatic: Slow Ramp, Fast Ramp, 25 % step

*Temperature range of 18 °C to 28 °C, for one year after calibration

General Specifications

Maximum voltage applied between
any jack and earth ground:
1000 V RMS
Storage temperature: -40 °C to 60 °C
Operating temperature: -20 °C to 55 °C
Temperature coefficient: 0.05 x (specified accuracy) per °C
(for temperatures < 18 °C or > 28 °C)
Relative humidity 95 % up to 30 °C;
75 % up to 40 °C; 45 % up to 50 °C;
35 % up to 55 °C
Vibration: Random, 2 g, 5-500 Hz
Shock: 1 meter drop test
Safety: Designed in accordance with EN61010, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92 Over-voltage Category III
Size: 50 mm H x 100 mm W x 203 mm L
(1.97 in H x 3.94 in W x 8.00 in L)
Weight: 600 g (1.3 lbs)
Battery: Four AA alkaline batteries
Battery life: 140 hours typical (measurement)
10 hours typical (sourcing 12 mA)
Warranty: 3 years