Specifications (18° to 28° C, one year)
Voltage
Measure
Range:
0
to 28 V
Resolution:
.001
V
Accuracy:
±(.015%
Rdg + 2 counts)
mA
measure
Range:
0
to 24 mA
Resolution:
.001
mA
Accuracy:
.015%
Rdg + 2 LSD
mA
source1
Range:
0
to 24 mA
Resolution:
.001
mA
Accuracy:
.015%
Rdg + 2 LSD
mA
simulate2
Range:
0
to 24 mA
Resolution:
.001
mA
Accuracy:
.015%
Rdg + 2 LSD
Loop
supply
Range:
24
V dc
Resolution:
n/a
Accuracy:
=
24 V +/- 10%
Temperature
Coefficient, -10 to 18 șC, 28 to 55 șC, ± .005% of range per șC;
1 Max load, 1200 Ohms; 950 Ohms at 20 mA in HART mode
2 Max applied voltage for simulation, 30 V
Operating
Modes
Measure
current
Measure
current/Loop power
Measure
voltage
Simulate
current; manual adjust; scroll up/down
Simulate
current; 25% steps; 0 - 100% steps
Simulate
current; Auto ramp (smooth/step)
Source
current; manual adjust; scroll up/down
Source
current; 25% steps; 0 - 100% steps
Source
current; Auto ramp (smooth/step)
General
Specifications
Maximum
voltage
30
volts
Non-operating
temperature
-30
to 60 șC
Operating
temperature
-10
to 55 șC
Relative
humidity
0
to 90% (0 to 35 șC);
0 to 70% (35 to 55 șC)
Operating
altitude
3,000
meters max.
Shock
and vibration
Per
MIL-T-28800 for a Class 2 Instrument
Safety
CSA
C22.2 No. 1010.1: 1992
EMC
EN50082-1:1992
and
EN55022:1994 Class B
Size/weight
6.0
in L x 2.8 in W x 1.6 in D
(152 mm L X 70 mm W x 41 mm D)
8 oz (224 g)
Size/weight
with C10 Holster
6.4
in L x 2.9 in W x 1.9 in D
(164 mm L x 75 mm W x 47 mm D)
12.3 oz. (350 g)
Battery:
One
9 V alkaline
Battery
life:
18
hours typical, at 12 mA
Warranty
Three
years