Bandwidth limiter: User selectable 10 kHz or
20 MHz
Number of inputs: 2 plus external trigger.
All isolated from each other and ground.
Input coupling: AC, DC with ground level
indicator
Input sensitivity: 5 mV/div to100 V/div
Input voltage: See general specifications
for maximum rating
Vertical resolution: 8 bit
Accuracy: ±(1.5% of reading + 0.04 x
range/div)
Input impedance: 1 MW ± 1% // 15 pF ± 2 pF
Horizontal
Maximum real-time sample rate:
Fluke 199= 2.5GS/s
Fluke 196=1 GS/s
Fluke 192=500 MS/s
Number of digitizers: 2
Time base range:
Fluke 199 and 196= 5 ns/div to 5 s/div
Fluke 192= 10 ns/div to 5 s/div
Maximum record length: 1,000 points per
input
27,500 points per input in ScopeRecord-roll mode
(10 ms/div ... 2 min/div)
Accuracy: ± (0.01% of reading + 1 pixel)
Glitch capture: 50 nsec (5 µsec/div to 1 min/div) Faster time bases have higher
sample rates than
20 MS/s
Source: Input A, input B, external trigger input. All
input references isolated from each other and ground.
Modes: Automatic Connect-and-View, free run, single shot, edge,
delay, video, video line, selectable pulse width
Connect-and-View: Advanced automatic
triggering that recognizes signal patterns, automatically sets up and continuously adjusts
triggering, time base and amplitude. Automatically displays stable waveforms of complex
and dynamic signals like motor drive and control signals.
Video triggering: NTSC, PAL, PAL+, SECAM.
Includes field 1 and 2 and line select.
Pulse width triggering: Pulse width
qualified by time. Allows for triggering t, = t, ą t, where t is selectable in minimal
steps of 0.01 div or 50 nsec.
Time delay: 9 divisions pre-trigger view to 1,000
divisions trigger delay
Automatic capture of 100 screens
The instrument ALWAYS memorizes last 100 screens (no user
interaction or setup required). When an anomaly occurs on screen, theres 10 seconds
to press HOLD and review it. If one sets up the instrument for triggering on glitches or
intermittent anomalies the unit operates in baby-sit mode and will capture 100
events.
Replay: Manual or continuous
replay. Displays the captured 100 screens as a live animation. The contents
can also be viewed by manually scrolling backwards and forwards screen by
screen.
Replay storage: Up to 2 sets of 100 screens can be
saved for later recall and analysis.
Automatic scope measurements
VDC, VAC rms, VAC+DC, Vpeak max, Vpeak min, Vpeak to peak,
frequency (Hz), positive pulse width, negative pulse width, positive duty cycle, negative
duty cycle, amp AC, amp DC, amp AC+DC, power factor, watts, VA, VA reactive, phase,
temperature °C, temperature °F, dBV, dBm into 50W and 600W
Cursor Measurements
Source: Input A or B
Dual horizontal lines: Voltage at cursor 1 and 2,
voltage between cursors
Dual vertical lines: Time between cursors,
voltage between markers
Single vertical line: Min-Max and Average voltage at cursor
position
Via 4 mm banana inputs. Fully isolated from scope inputs and
scope ground. The specified accuracy is valid over the temperature range 18 °C to 28 °C
(15 °F to 33 °F). Add 10 % of specified accuracy for each degree °C below 18 °C or
above 28 °C (15 °F to 33 °F).
Maximum resolution
5,000 counts
Voltmeter ranges
500mV, 5V, 50V, 500V, 1,000V
Accuracy
VDC: ± (0.5 % + 5 counts)
VAC true rms
15 Hz to 60 Hz: ± (1 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
VAC+DC true rms
DC to 60 Hz: ± (1 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
Design: Rugged, shock proof with integrated
protective holster
Drip and dust proof:
IP51 according to IEC529
Shock and vibration: Shock 30g, Vibration 3g
according to MIL-T28800E, Type III, Class 3, Style B
Display
Bright LCD with CCFL backlight, 35/60 cd/m2 without/with
power adapter.
Size: 105 x 79 mm (4.1
x 3.1 inches)
Resolution: 240 x 240 pixels
Contrast and brightness: User adjustable,
temperature compensated
Memory Save and Recall
Scope memories: 10 memory locations that each can
contain two waveforms plus corresponding setup.
Recorder memories: 2 memory locations that each can contain 100 captured
dual input scope screens, or a dual input ScopeRecord (27,500 Min-Max pairs per input), or
a dual input Trendplot (13,500 Min-Max pairs per input).
Real-time Clock
Time and date stamp for ScopeRecord, 100 captured screens and
Trendplots.
Power
Line power: Country specific line voltage
adapter/battery charger included
Battery
power: Rechargeable NiMH (installed)
Battery operating time: 4 hours
Battery charging time: 4 hours
Battery power saving functions: Auto power
down with adjustable power down time. On screen battery power indicator
Maximum probe voltage: 1,000V CAT II, 600V CAT III
(Maximum voltage between 10:1 probe tip (VP190) and reference lead)
Floating voltage: 1,000V CAT II, 600V CAT III
(Maximum voltage between earth ground and any terminal (signal input or shielding))
Independently isolated: 1,000V CAT II, 600V
CAT III
inputs (Maximum voltage between any terminal of
one input or probe (VP190) and any other terminal
of another input or probe (VP190))
Maximum voltage on BNC input directly
(input A or B): 300V CAT III
Maximum voltage on meter input: 1,000V CAT
II, 600V CAT III
Environmental
Operating temperature: 0 °C to +50 °C
Storage temperature: -20 °C to +60 °C
Humidity
10 °C to 30 °C: 95% RH non condensing
30 °C to 40 °C: 75% RH non condensing
40 °C to 50 °C: 45% RH non condensing
Maximum operating altitude: 3,000 m (10,000
feet)
Maximum storage altitude: 12 km (40,000
feet)
Electro-Magnetic Compatibility: EN 61326-1
for emission and immunity
Optically isolated PC/Printer interface
To printer: Supports HP Laserjet®,Deskjet®,Epson
FX/LQ® and postscript printers via optional PAC 91
To PC: Transfer instrument settings, screen images
and waveform data, compatible with FlukeView®software for Windows®
via optional PM9080.