Waveform Mathematics: A+B, A-B, A*B
all with user selectable scaling of the result ; A versus B (XY-mode)
Acquisition modes: Normal, auto, single
shot, ScopeRecord, roll, glitch capture
Trigger and delay
Source: Input A,
input B, external trigger input. All input references isolated from each other and ground.
Modes: Automatic
Connect-and-View, free run, single shot, edge, delay, video, video line, selectable
pulse width
Connect-and-View: Advanced automatic
triggering that recognizes signal patterns, automatically sets up and continuously adjusts
triggering, time base and amplitude. Automatically displays stable waveforms of complex
and dynamic signals like motor drive and control signals.
Video triggering: NTSC, PAL, PAL+, SECAM.
Includes field 1 and 2 and line select.
Pulse width triggering: Pulse width
qualified by time. Allows for triggering t, = t, ą t, where t is selectable in minimal
steps of 0.01 div or 50 nsec.
Time delay: 12 divisions pre-trigger view
to 1,000 divisions trigger delay
Automatic capture of
100 screens
The instrument ALWAYS
memorizes last 100 screens (no user interaction or setup required). When an anomaly occurs
on screen, theres 10 seconds to press HOLD and review it. If one sets up the
instrument for triggering on glitches or intermittent anomalies the unit operates in
baby-sit mode and will capture 100 events.
Replay: Manual or continuous replay.
Displays the captured 100 screens as a live animation. The contents can also
be viewed by manually scrolling backwards and forwards screen by screen.
Replay storage: Up to 2 sets of 100 screens
can be saved for later recall and analysis.
Automatic scope
measurements
VDC, VAC rms, VAC+DC,
Vpeak max, Vpeak min, Vpeak to peak, frequency (Hz), rise time with markers, fall time
with markers positive pulse width, negative pulse width, positive duty cycle, negative
duty cycle, amp AC, amp DC, amp AC+DC, power factor, watts, VA, VA reactive, phase,
temperature °C, temperature °F, dBV, dBm into 50 W and 600 W
Cursor Measurements
Source: Input A
or B
Dual horizontal lines: Voltage at
cursor 1 and 2, voltage between cursors
Dual vertical lines: Time between cursors,
voltage between markers. Automatic rise time with markers at 10% and 90%, fall time with
markers at 10% and 90%
Single vertical line: Min-Max and Average
voltage at cursor position
Via 4 mm banana inputs.
Fully isolated from scope inputs and scope ground. The specified accuracy is valid over
the temperature range 18 °C to 28 °C (15 °F to 82 °F). Add 10 % of specified accuracy
for each degree °C below 18 °C or above 28 °C (15 °F to 82 °F).
Design: Rugged,
shock proof with integrated
protective holster
Drip and dust proof: IP51 according
to IEC529
Shock and vibration: Shock 30 g, Vibration 3
g
according to MIL-T28800E, Type III, Class 3, Style B
Display
Bright LCD with CCFL
backlight, 50/80 Cd/m2 without/with power adapter.
Size: 115 x 86 mm (4.5 x 3.4 inches)
Resolution: 320 x 240 pixels
Contrast and brightness: User adjustable,
temperature compensated
Memory Save and Recall
Scope memories: 10
memory locations that each can contain two waveforms plus corresponding setup. Plus 5
extra memory locations after web refistration.
Recorder memories: 2 memory locations
that each can contain 100 captured dual input scope screens, or a dual input ScopeRecord
(27,900 Min-Max pairs per input), or a dual reading Trendplot (18,000 Min-Max pairs and
avg.
Real-time Clock
Time and date stamp for
ScopeRecord, 100 captured screens and Trendplots.
Power
Line power:
Country specific line voltage adapter/battery charger included
Battery power: Rechargeable NiMH
(installed)
Battery operating time: 4 hours
Battery charging time: 4 hours
Battery power saving functions: Auto power
down with adjustable power down time. On screen battery power indicator
Maximum probe voltage:
1,000 V CAT II, 600 V CAT III
(Maximum voltage between 10:1 probe tip (VPS200) and reference lead)
Floating voltage: 1,000 V CAT II, 600
V CAT III
(Maximum voltage between earth ground and any terminal (signal input or shielding))
Independently isolated: 1,000 V CAT II, 600
V CAT III
inputs (Maximum voltage between any terminal of
one input or probe (VPS200) and any other terminal
of another input or probe (VPS200))
Maximum voltage on BNC input directly
(input A or B): 300 V CAT III
Maximum voltage on meter input: 1,000 V CAT
II, 600 V CAT III
Environmental
Operating temperature:
0 °C to +50 °C
Storage temperature: -20 °C to +60
°C
Humidity
10 °C to 30 °C: 95% RH non condensing
30 °C to 40 °C: 75% RH non condensing
40 °C to 50 °C: 45% RH non condensing
Maximum operating altitude: 3,000 m (10,000
feet)
Maximum storage altitude: 12 km (40,000
feet)
Electro-Magnetic Compatibility: EN 61326-1
for emission and immunity
Optically isolated
PC/Printer interface
To printer:
Supports HP Laserjet®,Deskjet®,Epson FX/LQ® and
postscript printers via optional PAC 91
To
PC: Transfer instrument settings, screen images and waveform data, compatible with
FlukeView®software for Windows® via optional PM9080.